Solartron 1287A Electrochemical Interface
SOLD OUT sold
lcr119






Measurement Configuration;
Cell connections:2,3 or 4 Terminal,all floating,
Working electrde;
Current measurement resistor range:0.1ohm to 1Mohm,
Full scale current range:2A to 200nA,
Limit of error:0.1%+-0.05% of range,
Counter electode;
output voltage,wrtLO:>+-30V,
Current,subject to thermal protection limits:2A,
slew rate,potentiostatic control:>10V/us,
output short-circuit protected,
Referece electrodes;
input impedance:>10Gohm,
capacitance:50pF,
current:<1nA,
Limit of error:0.1%+-100uV,
rejection,f,10kHz:75dB,f<1MHz:40dB,

Polarisation;
DC polarisation;
voltage range:+-14.5V,
limits of error V<3.2V:0.2%+-200uV,V>3.2V:0.2%+-2mV,
maximum resolution:100uV,
current range:+-2A,
limit of error:0.2%+-0.1% of range,
maximum resolution:100pA,
DC sweep:analog ramp;
ramp rate(voltage):6mV/min to 6000V/min,
minimum segment duration:10ms,
maximum segment duration:10E5s,
DC sweep:stepped ramp;
minimum step heigt:5uV/5pA,
maximum step height:29V/4A,
minimum step duration:10ms,
maximum step duration:10E5,
AC input;
voltage range:+-10V,
gain:x1,x0.01,
impedance:10kohm,
Control Loop Bandwidth,100ohm resistive load,unity gain;
potentiostatic mode,type C stability:>1MHz,
galvanostatic mode,type B stability:>100kHz

sold Thank you

BNC-Juck cable*4,Power Code,E-manual,