Analog unit 8956*4 •2ch Voltage •Input connectors Isolated BNC connector (input impedance 1 Mohm //40 pF) Max. rated voltage to earth: 300 V AC, DC (with input isolated from the unit, the maximum voltage that can be applied between input channel and chassis and between input channels without damage) •Measurement range: 5 mV - 20 V/division, 12 ranges, full scale: 20 divisions, AC voltage for possible measurement / display using the memory function: 280 V rms low-pass filter: 5Hz/500Hz/5kHz/1MHz •Measurement resolution 1/100 of measurement range (using 12-bit A/D conversion; installed in 8860 series) •Highest sampling rate 20 MS/s (simultaneous sampling in 2 channels)
Number of input Units:Max 4 units Number of channels:Max. 16 analog channels Max. 128 channels with scanner unit) + 16 logic channels (standard configuration) Measurement ranges(20 div full-scale): Measurement functions: MEM (high-speed recording) REC (real-time recording) REC & MEM (realtime recording + high-speed recording) FFT (frequency analysis) Real-time Save (records directly to storage media) Data storage media:PC Card Type II slot × 2, Hard disk drive (option) × 1 Display:10.4 inch TFT color LCD (SVGA, 800 × 600 dots) External interfaces:GP-IB (PC card), USB2.0 (series A receptacle 3 ports), LAN (HTTP/FTP/ Send mail or other), Monitor output (15-pin D-sub, SVGA output) Option 9718-50(HDD unit), 9715-51:128 MW 8995:A4 printer Unit
probe(L9197)*8,8956 analog unit*4,CD-ROM(wave viewer Ver1.38,MR8904 CAN Editor Ver1.03,RS232C/GPIB/LAN Manual,USB Device Driver