Yokogawa AQ6319 600-1700nm Optical Spectrum Analyzer
JPY1,800,000 1 stock
opt1061







フロントパネル傷あり

Applicable fiber:SM (9.5/125 μm), GI (50/125 μm),
Measurement wavelength range:600 to 1700 nm,
Span:0.1nm to full range and zero span,
Wavelength Accuracy:+-10pm(1520 to 1580nm),+-20pm(1450 to 1520nm,1580 to 1620nm),+-50pm(Full Range),
Wavelength Linearlity:+-10pm(1520 to 1580nm),+-20pm(1450 to 1520nm,1580 to 1620nm),
Wavelength repeatability:+-2pm(<1min,1450 to 1620nm),
Measurement data point:101 to 50001,
Resolution Bandwidth:0.01,0.02,0.05,0.1,0.2,0.5 and 1nm,
Maximum input power:+23 dBm,
Level sensitivity:-90dBm(1250 to 1620nm),-80dBm(1000 to 1250nm),-60dBm(800 to 1000nm,1620 to 1680nm),
Level accuracy:+-0.3dB(1550/1600nm,0/-20dBm,RES:>0.02nm),+-0.3dB(1310nm,0/-20dBm,RES:>0.05nm),
Polarization dependency:+-0.05dB(1520 to 1620nm,RES:>0.02nm),
+-0.07dB(1450 to 1520nm,RES:>0.02nm),
+-0.07dB(typ).(1310nm,RES:>0.05nm),
Sweep Time:0.5sec(SENS:NORE_HOLD),1sec(SENS:MID),3sec(SENS:HIGH1),15sec(SENS:HIGH3),120sec(SENS:HIGH3 with chop mode on),
Analysis:LED, FP-LD, DFB-LD, WDM, EDFA,
Optical filter, WDM filter, Spectral width,
Notch width, SMSR, PMD, Power,Go/NoGO judgment,
Auto analysis, Analysis between linemarkers,
Analysis in the area,
Storage:FDD (3.5-inch 2HD),Internal storage,
File format
Interface:GP-IB, RS-232, Ethernet (TCP/IP),
Power requirement:100 to 240 (±10%) V, 50/60 Hz, approx. 400 VA,
Dimensions and mass:425 (W)  222 (H)  500 (D) mm, Approx. 33 kg,