Source Frequency Range:5 Hz to 3 GHz Resolution:1mHz Source stability:± 7 ppm (5 to 40 °C) CW accuracy:± 1 ppm ± 1 mHz
Level accuracy (Option 3L3/3L4/3L5) ± 0.8 dB (at 0 dBm, 50 MHz absolute) ± 1.0 dB (at 5 Hz to 1.5 GHz, 0 dBm, relative to 50 MHz) ± 1.5 dB (at 1.5 GHz to 3 GHz, 0 dBm, relative to 50 MHz)
Level lineality (Option 3L3/3L4/3L5) ± 0.75 dB (at –10 to 10 dBm, 0 dBm reference)
Output Power Range (Option 3L3/3L4/3L5) –45 dBm to 10 dBm Level resolution:0.05 dB
Input
Absolute amplitude accuracy < ± 3 dB (at 0 dBm)
Crosstalk (Option 3L3/3L4/3L5) –85 dB (at 5 Hz to 100 Hz) –100 dB (at 100 Hz to 9 kHz) –110 dB (at 9 k to 100 kHz) –115 dB (at 100 kHz to 3 GHz)
Noise floor (Option 3L3/3L4/3L5) 10 Hz bandwidth -90 dBm (100 Hz to 9 kHz) -100 dBm (9 kHz to 100 kHz) -105 dBm (100 kHz to 1 MHz) -110 dBm (1 MHz to 3 GHz)
E5061B-3L5 + 005 Impedance Analyzer function
Impedance parameters:IZI, θz , IYI, θy, Cp, Cs, Lp, Ls, Rp, Rs, D, Q, R, X, G, B
Measurement methods:Port 1 Reflection, Port 2 Reflection, Port 1-2 Series-thru, Port 1-2 Shunt-thru, Gain-phase Series-thru, Gain-phase Shunt-thru
Equivalent circuit analysis:3-component model (4 models: A/B/C/D), 4-component model (1 model: E)
Test frequency: 5 Hz to 500 MHz/1.5 GHz/3 GHz (Port-1 and 2), 5 Hz to 30 MHz (Gain-Phase test port), 1 mHz resolution AC source power(Osc level): –45 to + 10 dBm (when the source port is terminated with 50 Ω) DC voltage bias: 0 to ±40 V (100 mA max.) 1 mV resolution (0 V to ±10 V), 4 mV resolution (10 V to 40 V, -10 V to -40 V), DC bias monitor available (voltage or current)
Option: 3L5 LF-RF network analyzer with DC bias source, 5 Hz to 3 GHz 1E5 High stability timebase 020 Standard Data Storage 005 Impedance analysis for LF-RF network analyzer